OCT Imaging Set-Up & Prep

Horizontales
True or False: You do not need to make sure that OCT is selected on the StarTech box (Image I) to display the OCT screen on the boom.
The guide catheter should be ____ French with no side holes for cases using OCT.
If OCT is going to be performed in a case using a manifold and not the power injector, make sure to grab a ____ cc control syringe from the Extra OCT Parts bin located to the left of the OCT catheters. Fill the control syringe with at least 15ccs of 100% contrast.
The purpose of purging the Dragonfly imaging catheter outside of the body with the 3cc control syringe is to remove ___ from the catheter.
To turn the OCT system ____, click the button shown in Image C.
The name of the setting on the Medrad power injector that needs to be selected for cases using OCT.
If you see an Extreme Ring artifact (Image H) during a case using OCT, grab another Dragonfly imaging _____. This dims the image and can cause calibration failures. This is usually due to a partial fiber break. NOTE: Important to avoid kinking or bending the imaging catheter while handling throughout the procedure to prevent this.
Verticales
Remove the Dragonfly imaging catheter from the hoop, wipe shaft with heparinized saline to activate the hydrophilic coating, fill 3cc control syringe with ___ % contrast, remove all air bubbles from syringe, and purge catheter until you see at least 3 drops exit the distal tip (prior to connecting to the DOC controller). Make sure to leave the 3cc control syringe connected throughout the case for repeat purging.
On the angio co-registration, place the _____ point first just distal to the guide catheter tip and then place the distal point on the guidewire tip (Image K).
The label on the bin that stores extra DOC covers, 3cc control syringes, and 20cc control syringes in the storage closet in between Lab 4 and Lab 5 (Image J). This bin is to the left of the OCT catheters.
To turn off the OCT system, click the ____ button in the top right corner of any screen (Image A) and select logout. Then from the login screen click the power button in the bottom right corner (Image B) to shut down.
True or False: If you see the error shown in Image D after inputting the username and password into the OCT system click “ok” and proceed using OCT in the case. This error does not impact the performance of the system.
When entering in the patient's DOB it is important to click the _____ mark (Image F) before clicking the Add button that submits the patient's info.
True or False: These are the steps to start an OCT case on the monitor in the control room (Image E): (1) Click begin a new case (2) Click add new patient (3) Enter in patient info. Important: Make sure to click the check mark that pops up under the DOB to save the DOB (4) Select the physician’s name (5) Click start OCT (6) Select 100% contrast, vessel, and location.
True or False: The button on the Carot monitoring system that displays OCT on the boom is either listed as CoroFlow/OCT or OCT/CoroFlow (Image G)